Traymorrow · Sterile processing learning guide

Cleaning and Inspecting Reusable Devices Before Sterilization

A source-linked learning guide for healthcare facilities, surgical centers, and sterile processing teams. Follow the full device-specific reprocessing process from point-of-use handling through cleaning, inspection, packaging, sterilization, and storage.

All sterile processing resources

One checkpoint cannot replace the sequence.

Cleaning and inspection sit inside a larger reprocessing sequence. FDA describes initial point-of-use processing, transfer for thorough cleaning, then the device-appropriate disinfection or sterilization and return to storage or use. CDC notes that the sterile result also depends on decontamination, preparation, packaging, loading, monitoring, and choosing the appropriate cycle. [1][3]

No single cleaning, visual inspection, or sterilizer step can stand in for the complete process that applies to a device. Cleaning removes soil that can interfere with later processing; inspection is a checkpoint, not sterilization. Follow your facility’s current procedures and the applicable instructions for use (IFUs) for the device, cleaning products, washer, packaging system, and sterilizer at each relevant step. [1][2][3][4]

From point of use to protected storage

  1. 1. Start at the point of use

    Begin the device-specific point-of-use steps promptly after use. CDC recommends cleaning as soon as practical because soil can dry onto instruments, making it harder to remove and making later disinfection or sterilization less effective. Use only the treatment, moisture-management method, and handling steps specified by the device IFU and facility procedure; this guide does not set a universal method or time. [1][2]

  2. 2. Transport contaminated items to processing

    Move used items through the facility’s designated route to the appropriate decontamination or reprocessing area. Keep the contaminated status clear and follow facility procedures for containment, worker protection, and safe handling. FDA describes transfer from point of use to the reprocessing work area as part of the sequence; CDC advises treating used items sent to central processing as contaminated unless they were decontaminated at the point of origin. [1][3]

  3. 3. Decontaminate and clean to the device instructions

    Thorough cleaning comes before sterilization. CDC recommends meticulous cleaning to remove visible organic residue and inorganic salts, using a compatible detergent or enzymatic cleaner and a manual or mechanical method suited to the item. Follow the device and equipment IFUs for disassembly, agents, equipment, rinsing, and any device-specific steps. FDA emphasizes that each reusable device needs its own reprocessing regimen. [1][2][4]

  4. 4. Inspect after cleaning, before preparation

    Inspect the cleaned device at the point required by the applicable IFUs and facility procedure. Check for remaining visible soil and for damage or breaks in integrity that could interfere with cleaning or later processing. CDC recommends inspecting equipment surfaces for breaks in integrity; a CDC infection-prevention pocket guide also calls for visually checking reusable devices for residual soil before disinfection or sterilization. If a check fails, follow the facility’s documented re-cleaning, repair, or removal-from-service process before proceeding. [2][5]

  5. 5. Prepare and package clean devices

    After cleaning, drying, and inspection, prepare the item and package or container for the intended process. CDC describes the packaging area as the place to inspect, assemble, and package clean material, and says items requiring sterilization should be wrapped or placed in an appropriate rigid container. Use the device, packaging-system, and sterilizer instructions for preparation and compatibility. [3][4]

  6. 6. Run the applicable sterilization process

    When the device’s intended use requires sterilization, use the validated cycle and load configuration applicable to that device and the equipment in use. CDC recommends following the manufacturers’ sterilization parameters for the instrument, sterilizer, and container or wrap, and using process monitoring. Do not use a sterilizer cycle to compensate for incomplete cleaning or inspection. [2][3]

  7. 7. Release, store, and route the item safely

    Complete the facility’s load review and release steps, then protect the processed item during storage and transport so its package remains intact. CDC recommends protecting sterile items from conditions that compromise packaging, and reprocessing a pack before use when its integrity is compromised. Follow the facility policy and the applicable storage and transport instructions for the device and packaging system. [1][2]

References

Citations identify the organization, source title, and direct source link. The cited CDC pages present recommendations; the FDA 2015 guidance describes itself as recommendations for industry and FDA staff. None of these links replaces the facility’s applicable requirements or device-specific instructions.

  1. [1]

    U.S. Food and Drug Administration (FDA). How are Reusable Medical Devices Reprocessed?

    FDA overview of the general sequence: point-of-use processing, thorough cleaning, then device-appropriate disinfection or sterilization and storage or return to use.

    Open source
  2. [2]

    Centers for Disease Control and Prevention (CDC). Recommendations for Disinfection and Sterilization in Healthcare Facilities

    CDC guideline recommendations with evidence categories. They are recommendations, not a universal facility-specific policy; applicability depends on the relevant device and requirements.

    Open source
  3. [3]

    Centers for Disease Control and Prevention (CDC). Sterilizing Practices — Guideline for Disinfection and Sterilization in Healthcare Facilities (2008)

    CDC guidance describing cleaning, preparation, packaging, sterilization, and storage as connected parts of the process; the page identifies its source guideline as 2008.

    Open source
  4. [4]

    U.S. Food and Drug Administration (FDA). Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling — Guidance for Industry and Food and Drug Administration Staff (2015)

    Final FDA guidance about manufacturers’ reprocessing instructions and validation. FDA describes the document as recommendations; it is not a substitute for device-specific IFUs or facility procedures.

    Open source
  5. [5]

    Centers for Disease Control and Prevention (CDC). Preventing Patient Infections: My Pocket Guide

    CDC quick-reference guidance for orthopedic and pain procedures; it includes visual inspection for residual soil before disinfection or sterilization. It is cited as a recommendation, not as a universal rule for every device.

    Open source